For process water
Range: 0.5 to 200 ppm
Output signal: 4-20 mA
Pressure limit: 1 bar as a maximum
Unnecessary zero adjustment
DESCRIPTION
Principle Free chlorine measurement by amperometric method with a diaphragm cell of 2 electrodes; wasted sample. The probe includes a CTN sensor for the temperature compensation.
Mounting / Recommendations The measuring at a constant flow rate requires the use of a specific cell (see data sheet 193-95). The complete assembly optimizes the operations.
CAUTION
The sample may not content surfactants and the pH must be lower than 7.4
TECHNICAL FEATURES
Range: 0.5 to 200 ppm (free chlorine) Measuring signal depends of pH; regulation of pH is necessary.
Operating pH range: From 4 to 7.4 and regulated Pressure limit: 1 bar as a maximum Temperature limits: From 1 to 45°C Flow rate limits: From 30 to 40 L/h (see the data sheet 193-95) Power supply: 12 ... 30 V DC, [Rmax = (U-7,5) / 20 kOhm] Materials: PVC-U Dimensions: Diam.25 mm, length 225 mm